Saniscope Solution Test Strips’ are the most reliable tool for testing and monitoring the minimum effective concentration (MEC) of Saniscope solution during its use life
Saniscope Solution Test Strips should only be used with Saniscope solution before each usage to guard against dilution, which may lower the peracetic acid level of the solution below its MEC. Before immersing instruments, test the solution.
Completely dip indicating pad of the strip into Saniscope solution. Hold for two seconds and remove. Do not shake the strips after removal. Remove excess solution by holding the strip upright on a paper towel. Read the results at 90 seconds after removal. The intensity and colour shade of the pad should be equal or more than the colour shade displayed as ‘PASS 1’ in the below chart in order to indicate that Saniscope solution is effective.

If read past 90 seconds color may gradually change to indicate false results.